Close Menu
  • News
  • Medical
  • Technology
  • Nanomaterials
  • Research
  • Blog
    • Nasiol.com
  • Contact
    • Tech7685@gmail.com
What's Hot

Unique molecule may lead to smaller, more efficient computers

May 9, 2025

Depositing quantum dots on corrugated chips improves photodetector capabilities

May 8, 2025

Ultrasonic nanocrystal surface modification restores stainless steel’s corrosion resistance

May 7, 2025
Facebook X (Twitter) Instagram
Nanotech – Nanomaterials | Medical | Research | News Stories Updated Daily Nanotech – Nanomaterials | Medical | Research | News Stories Updated Daily
  • News
  • Medical
  • Technology
  • Nanomaterials
  • Research
  • Blog
    • Nasiol.com
  • Contact
    • Tech7685@gmail.com
Facebook X (Twitter) Instagram
Nanotech – Nanomaterials | Medical | Research | News Stories Updated Daily Nanotech – Nanomaterials | Medical | Research | News Stories Updated Daily
Home»News»Optimizing Particle Size Analysis for Light-Absorbing Colloidal Suspensions with the BeNano 180 Zeta Pro
News

Optimizing Particle Size Analysis for Light-Absorbing Colloidal Suspensions with the BeNano 180 Zeta Pro

April 14, 2024No Comments4 Mins Read
Facebook Twitter Pinterest Telegram LinkedIn Tumblr WhatsApp Email
Optimizing Particle Size Analysis for Light-Absorbing Colloidal Suspensions with the BeNano 180 Zeta Pro
Share
Facebook Twitter LinkedIn Pinterest Telegram Email

Darker-colored samples commonly demonstrate significant light absorption, resulting in diminished intensity and fluctuations in scattered light during dynamic light scattering tests. As a result, it becomes imperative to dilute the sample to ensure precise results.

This article uses the BeNano 180 Zeta Pro to analyze the particle size of two samples in aqueous solutions with light-absorbing effects.

Instrumentation

The BeNano 180 Zeta Pro features a 671 nm solid-state laser with a power of 50 mW. Dynamic light scattering (DLS) is performed by collecting scattered light signals from the sample at 90° or 173°. The calculation of intensity fluctuation yields the correlation function from which the diffusion coefficient can be derived. Utilizing the Stokes-Einstein equation, the size and size distribution of the sample are determined.

Experiment

Both samples in this investigation showed brown coloration, indicating strong light absorption. Particle size analysis was performed on two samples under various circumstances, including stock solutions and subsequent dilutions (2×, 10×, and 100×).

All measurements were taken at 25 °C ±0.1 °C using the 90° and 173° optics. The repeatability of the results was assessed by measuring each sample three times.

Results and Discussion

Correlation functions (top) and size distributions (bottom) of 2× diluted sample #1 detected at 90°.

Figure 1. Correlation functions (top) and size distributions (bottom) of 2× diluted sample #1 detected at 90°. Image Credit: Bettersize Instruments Ltd.

Correlation functions (top) and size distributions (bottom) of 2× diluted sample #2 detected at 90°.

Correlation functions (top) and size distributions (bottom) of 2× diluted sample #2 detected at 90°.

Figure 2. Correlation functions (top) and size distributions (bottom) of 2× diluted sample #2 detected at 90°. Image Credit: Bettersize Instruments Ltd.

Correlation functions (top) and size distributions (bottom) of 2× diluted sample #1 detected at 173°.

Correlation functions (top) and size distributions (bottom) of 2× diluted sample #1 detected at 173°.

Figure 3. Correlation functions (top) and size distributions (bottom) of 2× diluted sample #1 detected at 173°. Image Credit: Bettersize Instruments Ltd.

Correlation functions (top) and size distributions (bottom) of 2× diluted sample #2 detected at 173°.

Correlation functions (top) and size distributions (bottom) of 2× diluted sample #2 detected at 173°.

Figure 4. Correlation functions (top) and size distributions (bottom) of 2× diluted sample #2 detected at 173°. Image Credit: Bettersize Instruments Ltd.

Figures 1–4 show consistent test results for both samples at varying concentrations and angles. The stock solution cannot be observed at a 90° optics angle due to the sample’s high absorption.

See also  Collaborative review unveils the potential of graphene in advancing nitride semiconductor technology

Table 1. Size results of sample #1. Source: Bettersize Instruments Ltd.

Sample #1 90° 173°
Dilution Factor Z-average Size Z-average Size
0 — 45.90 ± 0.23
2 29.74 ± 0.4 28.50 ± 0.12
10 26.11 ± 0.91 25.61 ± 0.08
100 25.86 ± 0.23 25.71 ± 0.11

 

Table 2. Size results of sample #2. Source: Bettersize Instruments Ltd.

Sample #2 90° 173°
Dilution Factor Z-average Size Z-average Size
0 — 34.38 ± 0.1
2 24.21 ± 0.44 24.93 ± 0.24
10 22.68 ± 0.2 22.98 ± 0.26
100 23.18 ± 0.24 22.55 ± 0.16

 

Z-average sizes of sample #1 (top) and #2 (bottom) obtained at different dilution ratios and at different detection angles.

Z-average sizes of sample #1 (top) and #2 (bottom) obtained at different dilution ratios and at different detection angles.

Figure 5. Z-average sizes of sample #1 (top) and #2 (bottom) obtained at different dilution ratios and at different detection angles. Image Credit: Bettersize Instruments Ltd.

Figure 5 shows a concentration-dependent reduction in size when measured at 90° and 173°. After a tenfold dilution, there are no additional size changes with concentration. The findings at 90° and 173° show consistent agreement under dilution conditions.

The size increases with concentration, primarily due to the sample’s pronounced absorption of scattered light. This absorption diminishes the intensity and fluctuations in scattered light, thus yielding larger results. Consequently, dilution is indispensable for samples displaying substantial absorption. However, within the dilution range of 10 to 100 times, the particle size remains stable, indicating minimal absorption of scattered light and thereby ensuring the accuracy of the size results.

Conclusion

Utilizing different detection angles, specifically 173° and 90°, revealed that the results obtained at 173° were not closer to the true value at higher concentrations. This is due to the fact that backscattering (173°) has the benefit of suppressing multiple light scattering for substances with large concentrations, such as milky white samples.

See also  Understanding the forces that regulate crystallization by particle attachment

Backscattering, however, is unable to improve detection capabilities at high concentrations for samples with substantial absorption. Therefore, rather than depending exclusively on instrument configurations, users are recommended to take their sample characteristics into account when choosing instrument specifications and detection angles.

This information has been sourced, reviewed and adapted from materials provided by Bettersize Instruments Ltd.

For more information on this source, please visit Bettersize Instruments Ltd.

Source link

Analysis BeNano Colloidal LightAbsorbing Optimizing Particle Pro Size Suspensions Zeta
Share. Facebook Twitter Pinterest LinkedIn Tumblr Email

Related Posts

Unique molecule may lead to smaller, more efficient computers

May 9, 2025

Depositing quantum dots on corrugated chips improves photodetector capabilities

May 8, 2025

Ultrasonic nanocrystal surface modification restores stainless steel’s corrosion resistance

May 7, 2025

Structure dictates effectiveness and safety in nanomedicine, driving therapeutic innovation, say scientists

May 7, 2025

How Can Nanomaterials Be Programmed for Different Applications?

May 6, 2025

Ultrafast plasmon-enhanced magnetic bit switching at the nanoscale

May 6, 2025

Comments are closed.

Top Articles
News

Researchers develop proton barrier films using pore-free graphene oxide

News

Researchers achieve >99% photoluminescence quantum yield in metal nanoclusters

News

Scientists develop dual-functional, high-efficiency antimicrobial nanozyme

Editors Picks

Unique molecule may lead to smaller, more efficient computers

May 9, 2025

Depositing quantum dots on corrugated chips improves photodetector capabilities

May 8, 2025

Ultrasonic nanocrystal surface modification restores stainless steel’s corrosion resistance

May 7, 2025

Structure dictates effectiveness and safety in nanomedicine, driving therapeutic innovation, say scientists

May 7, 2025
About Us
About Us

Your go-to source for the latest nanotechnology breakthroughs. Explore innovations, applications, and implications shaping the future at the molecular level. Stay informed, embrace the nano-revolution.

We're accepting new partnerships right now.

Facebook X (Twitter) Instagram Pinterest
Our Picks

Image-guided cancer treatment using biocompatible copper-based nanomaterials

August 6, 2024

Applications of Liquid Cell TEM

September 14, 2023

Creating vortices in a superfluid made of light

December 1, 2023

Subscribe to Updates

Get the latest creative Nano Tech news from Elnano.com

© 2025 Elnano.com - All rights reserved.
  • Contact
  • Privacy Policy
  • Terms & Conditions

Type above and press Enter to search. Press Esc to cancel.

Cleantalk Pixel